Products Description

Automatic Chemistry Biochemistry Analyzer is used to quantitatively analyze the clinical chemistry of human serum, plasma, urine, cerebrospinal fluid and other samples.
Automatic Chemistry Biochemistry Analyzer's Features
1. 37 Sample positions.
2. 28 Reagent positions.
3. 48 Reaction cuvettes.
4. 200 Tests per hour.
5. Probe anti-collision function,liquid level dectection function.


Automatic Chemistry Biochemistry Analyzer Operation steps
1.Prepare a new stirrer to be replaced.
2.Turn off the power of the analyzer and open the top cover of the analyzer.
3.Remove the reaction tray cover.
4.Rotate the maxing arm to the direction of the reaction tank by hand (the methodis like the sample arm operation)
Figure 13-41 Rotating stirring arm
5.Use a Phillips screwdriver to loosen the two fastening screws for one week.
6.Wipe the tip of the new stirrer with a cotton swab dipped in alkaline cleaningsolution,and then wipe the surface of the stirer with a cotton cloth dampenedwith pure water (do not bend the stirrer when wiping).
7.When installing the new stirrer,insert the stirrer into the root of the motor shaftand fix it with the M2 screw
Automatic Chemistry Biochemistry Analyzer
Safety Precautions
For use of this instrument safely,please read the following safety precautionscarefully.Any operation that violates the following safety precautions mayresult in personal injury or damage to the instrument.
Biological infection risk:
Improper use of the sample may result in infection.Do not touchsamples,controls,calibrators,mixtures,and waste directly with yourhands.Always wear gloves,work clothes to prevent infection,and wearprotective glasses if necessary.
If the sample is inadvertently in contact with the skin,please follow theuser's work standards immediately and consult a doctor.
Prevent getting involved in moving parts during operation
Caution of surface:
Do not touch the light source after the system is turned on.
When replacing the halogen lamp,the lamp must be replaced afterthe power is turned off and wait at least 20 minutes for the halogenlamp to cool down,otherwise the high temperature halogen lampand the light source box may cause burns.
Security information:
When working with the instrument,please do not look directly atthe light source or the light beam emitted by the barcode scanner.These beams will cause eye damage.
Before checking the light source,disconnect the analyzer from themains and wait at least 20 minutes until the light source cools.Donot touch the light source before it cools down to avoid burns.

Meter's Specifiction
| Production Name | Automatic Chemistry Biochemistry Analyzer | |
| Overall Performance | Model | BK-200 | 
| Throughput | 200 Tests/hour | |
| Analysis Method | End-point, Fixed-time, Rate(Kinetic) | |
| Sample & Reagent System | Sample Positions | 37 sample positions | 
| Reagent Position | 28 reagent positions | |
| Sample Volume | 2~70 | |
| Reagent Volume | 20~350 | |
| Probe | Liquid level sensor function, anti-collision function. | |
| Probe Washing | Automatic washing interior and exterior | |
| Reagent Cooling | Refrigerated tray with independent switch | |
| Calibration & QC | Calibration | Linear:K factor, 1-point ,2-point and multipoint liner Non-Linear:Spline,Polygon,Index,Ogarithm,Logit-4P, Logit-5P | 
| Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | |
| Data Management | Software | Windows 7/8/10 , 32or 64bit | 
| LIS System | Bi-direction,support HL7 protocol | |
| Interface | LAN port access | |
| Printer | External,multiple reporting mode available | |
| Size & Weight | Size (W*D*H) | 625*425*460mm | 
| Net Weight | 36kg | |
| Package Size (W*D*H) | 721*531*855mm | |
| Gross Weight | 50kg | |